Commissioning
44
MAX
MIN
RD
GN
1.
>1 s
2.
3.
A
Once the test button has been pressed, the current is 0.6 to 1 mA for at least 10 seconds, also if the button is
pressed for < 10 seconds. If the button is pressed for > 10 seconds, the current remains at 0.6 to 1 mA until
the test button is released. After this, the current is 2.2 to 3.5 mA again.
Pay attention to the information in the Safety Manual for proof-testing in safety
equipment in accordance with SIL.
8.7

Protecting settings from unauthorized access

8.8

Bluetooth parameters for Heartbeat

8.9

Verification wizard for proof testing

YE
2.2 ... 3.5 mA
+
2
1
0.6 ... 1 mA
+
2
1
A
+
2
1
Liquiphant FTL51B
RD
GN
YE
0.6 ... 1 mA
+
2
0.6 ... 1 mA
+
2
0.6 ... 1 mA
+
2
Endress+Hauser
1
1
1
A0033543