Electrical connection
36
6.2.10

Relay output

• Universal current connection with relay output (FEL64):
• Load AC: 253 V AC 6A (Ex de 4A)
• Load DC: 125 V DC 0.2A 30 V DC 6A (Ex de 4A)
• Pmax AC: 1500 VA, cos phi = 1 750 VA, cos phi > 0.7
• Direct current connection, relay output (FEL64DC):
• Load AC: 253 V AC 6A (Ex de 4A)
• Load DC: 125 V DC 0.2A 30 V DC 6A (Ex de 4A)
• Pmax AC: 1500 VA, cos phi = 1 750 VA, cos phi > 0.7
6.2.11

Cable entry

Required tools
For terminals: flat-blade screwdriver (0.6 x 3.5 mm)
1
1
Electronic insert
2
Connector M20x1.5
3
Maximum cable diameter 2.5 mm
4
Maximum cable diameter 4.0 mm (AWG12)
6.3

Special connection instructions

6.4

Hardware settings

6.5

Ensuring the degree of protection

6.6

Post-connection check

Is the device or cable undamaged (visual check)?
Do the cables comply with the requirements?
Do the cables have adequate strain relief?
Are the cable glands mounted and firmly tightened?
Does the supply voltage match the specifications on the nameplate?
If supply voltage is present, is the green LED lit?
2
3
4
ød
2
(AWG14)
Liquiphant FTL51B
3 mm
A0018023
Endress+Hauser