Electrostatic
± 8 kV
discharge
/Contact
(ESD) IEC
± 2 kV, ± 4
61000-4-2
kV, ± 8 kV,
± 15 kV /air
Electrical fast
± 2 kV for
transient/
power
burst IEC
supply lines
61000-4-4
± 1 kV for
input/outpu
t lines
Surge IEC
± 0.5 kV, ± 1
61000-
kV line(s) to
4-5
line(s)
± 0.5 kV, ± 1
kV, ± 2 kV
line(s) to
earth
Voltage dips,
<5% UT
short
(>95% dip in
interruptions
UT) for
and
0,5 cycle
voltage
40% UT
variations on
(60% dip in
power supply
UT) for 5
input
cycles 70 %
lines IEC
UT (30% dip
61000-4-11
in UT) for
25 cycles
<5% UT
(>95% dip in
UT) for 5 s
Power
30 A/m
frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
Radiated RF
Table 9 in
IEC 61000-4-3
IEC-60601-
1-2 2014
± 8 kV
Surface should be wood,
/Contact
concrete, or ceramic tile. If floors
± 2 kV, ± 4
are covered
kV, ± 8 kV, ±
with synthetic material, the
15 kV /air
relative humidity should be at
least 30%.
± 2 kV for
Mains power quality should be
power
that of a typical professional
supply lines
healthcare facility environment.
± 1 kV for
input/outpu
t lines
± 0.5 kV, ± 1
Mains power quality should be
kV line(s) to
that of a typical professional
line(s)
healthcare facility environment.
± 0.5 kV, ± 1
kV, ± 2 kV
line(s) to
earth
<5% UT
Mains power quality should be
(>95% dip in
that
UT) for
of a typical professional
0,5 cycle
healthcare facility environment.
40% UT
(60% dip in
UT) for 5
cycles 70 %
UT (30% dip
in UT) for
25 cycles
<5% UT
(>95% dip in
UT) for 5 s
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a
location in a typical professional
healthcare facility environment.
Table 9 in
Portable and mobile RF
IEC-60601-
communications equipment
1-2 2014
should be used no closer to any
Page 75