Compact Vital Signs Monitor
Procedural & Error Alarm Code Table continued
Alarm
LED
LCD Display
Code
Display
No
Low Battery
change
Flashing
battery icon
Blank
No code
Low Battery -
System
No
Disabled
change
Printer - No
Blank
Paper
NIBP Range
error
Other:
Blank
Error code,
N,P,E,I,S.
description
76
Audio Tone &
Effect of
Volume
Alarm
Silence
switch
3 beeps every
2 minutes
10 sec,
silence
adjustable
volume.
No effect
Steady tone,
maximum
2 minute
volume
silence
High priority
2 minute
alarm
silence
Volume
adjustable
High priority
alarm
Volume
adjustable
Steady tone,
No effect
maximum
volume
Effect of
Probable Cause
Clear via
Rotor
No effect
Replace or recharge battery. From
onset of alarm, 5 NIBP
measurements available. Beep
No effect
linearly increases rate as battery
discharges.
Clear
Replace or recharge battery. NIBP
measurement disabled.
Clear
Paper expired or printer door open.
NIBP algorithm returned value
outside specified accuracy rage.
No effect
Internal system fault

Advertisement