3.4 Monitor Disassembly

Most alignment and repair operations can only be performed with the monitor disassembled. This section
describes how to remove the covers, disassemble the chassis and remove the circuit boards. The
description refers to fully equipped DINAMAP
Compact TS models. DINAMAP
Compact T models do not
have SpO
functions and therefore the SpO
related components are not fitted.
2
2
Before beginning disassembly, disconnect the line power cord from the power socket and remove the
rechargeable lead-acid battery. Be careful to observe anti-static handling precautions throughout.
Note:
Pictures shown are from a revision 06 unit, appearance may differ slightly on earlier units.

3.4.1 Removal of the Front Panel

The front panel contains the Display board, Liquid Crystal Display module and switchboard. The Rotor,
Temperature probe socket and SpO
socket (model TS) are also mounted on the inside of the panel.
2
The front panel must be removed if access to any of the boards or modules mounted on it or behind it is
required. To remove the panel, release the two Pozidriv
screws mounted inside the hand recess of the
instrument' s rear enclosure.
Front panel
retaining screws
Then carefully pull the top of the front panel away from the rear enclosure. The panel will tilt forwards but is
still retained by two plastic clips at its bottom edge. Unhook the front panel from these clips, taking care not
to damage them. This is particularly important when refitting the front panel later.
42