Sample/Carrier Gas Supplies

Specification
Inlet Fitting
Inlet Pressure Range
Inlet Pressure Stability
Vent (Return) Pressure
Flow Rate
Temperature
Maximum Impurity Levels
* 1/4-inch face seal x 1/16-inch adapter fittings are available from ΑΜΕΤΕΚ.

Gas Ports

Specification
Carrier/Sample Inlet
Aux (Calibration)
Sample Outlet

Actuator Gas

Actuator Gas
Specification
Inlet Fitting
Inlet Pressure Range
Gas Type
Gas Type
(for ta7000R-O

Calibration Gas

Specification
Inlet Fitting
Cylinder Concentration
Cylinder Volume
Supply Pressure
Consumption
5-2 | ta7000R Gas Purity Monitor
Description
1/4-inch male face seal fitting*
4.8–6.1 BAR (70–90 PSIG)
±2 %, UHP regulator required
Atmospheric pressure vent is optimal, ±0.5 PSIG maximum
300 cc/minute minimum, sample bypass at 50 cc/minute
16 °C to 38 °C (60 °F to 100 °F), optimum when maintained ±2 °C
200 PPB Moisture, CO
Description
1/4-inch VCR compatible
1/16-inch VICI compression
1/16-inch VICI compression
1/8-inch VICI compression
Description
1/8-inch VICI compression fitting
4.8–10.2 BAR (70–150 PSIG)
Inerts (Nitrogen, Argon, Helium), 7 nines or better or CDA,
dewpoint < -65 °C
Inerts (Nitrogen, Argon, Helium), 7 nines or better
only)
2
Description
1/16-inch VICI compression fitting
5–10 PPM H
, CO, balance Nitrogen
2
100 cubic feet minimum (2800 L)
5.4–6.1 BAR (80–90 PSIG)
Approximately 1.0 L per calibration (4–8 L/month), automatic
shut-off between calibration.
, Hydrocarbons
2
PN T900-013, Rev A