Installing the Mechanical Components
Refer to Figures 2-1, 2-2, and 2-3 while installing the ta7000R Monitor.
The analyzer performance level depends upon the care and attention to
proper handling procedures during gas connections. Exposure to non-UHP en-
vironments or physical contact with sealing surfaces will reduce the accuracy
of the analyzers.
2-8 | ta7000R Gas Purity Monitor
The installation instructions must be strictly followed for safe and reli-
able operation of the ta7000R.
If any inconsistencies are encountered during installation, stop im-
mediately. Do not continue the installation without direction from
AMETEK Customer Service.
The ta7000R performance level depends upon the care and atten-
tion to proper handling procedures while making gas connections.
Exposure to non-UHP environments or physical contact with sealing
surfaces will reduce the accuracy of the ta7000R.
The vent must not cause any back-pressure in the ta7000R. Back-pres-
sure can damage the ta7000R.
PN T900-013, Rev A