Connecting Span Gas
The ta7000R Monitor uses Span gas for calibration purposes. The instrument
connection is made at the "AUX" 1/16-inch VICI port on the ta7000R Rear Panel.
Pressurize this external line to 80 PSIG (minimum 80 PSIG, maximum 90 PSIG).
Making the Vent Connections
All vent ports must be connected to a vent that is connected to a safe area. The
vent should be at atmospheric pressure.
2-12 | ta7000R Gas Purity Monitor
PN T900-013, Rev A