Ta7000R Monitor Location Requirements - Ametek ta7000R User Manual

Gas purity monitor
Hide thumbs Also See for ta7000R:
Table of Contents

Advertisement

Check the interior of the shipping container for the following items:

ta7000R Monitor Location Requirements

Observe the following guidelines when selecting an installation location:
2-4 | ta7000R Gas Purity Monitor
ta7000R Monitor
Power Cord
User Manual
Select a readily accessible position for the ta7000R to allow suitable access
to its Rear Panel for easy connections of the gases, and for routine service
and maintenance. Allow approximately 30" for service clearance at the
rear of the ta7000R. Comfort levels for maintenance personnel should be
considered in placement of the ta7000R and Display User Interface.
The installation location should:
-
Be free from excessive vibrations that can affect the electronics and
detectors.
-
Be within the ambient temperature limits listed in the specifications.
If the ambient temperature is outside the specified limits or the
vibration is excessive, contact your AMETEK representative to discuss
solutions and special options to address ambient temperatures.
-
Be clean with minimum particulate matter and dust in the air.
-
Be clear of any draft directly on the instrument from air conditioners
or extreme temperature fluctuations in the room.
Support gases for the instrument should be located near the ta7000R. A
uniform temperature environment for cylinders and regulators is essential
to maintain a constant pressure and flow rate. Always keep gas containers
inside the building where temperature fluctuations are at a minimum.
Have adequate ventilation around the instrument area without any high
flow velocity draft directly onto the ta7000R. Be sure not to block any vent
slots in the rear or top of the ta7000R since limited air circulation will cause
overheating and poor performance.
All connections to the instrument should be made with 1/16-inch stainless
steel tubing (maximum 0.030-inch ID). Coil some extra 1/16-inch gas lines
behind the ta7000R. This allows flexibility when moving or turning the
instrument for service and routine maintenance. Supplied by customer, or
can be purchased from AMETEK.
PN T900-013, Rev A

Advertisement

Table of Contents
loading

Table of Contents