Working From the Calibration Screen

From the Temperatures & Status screen press the Page key to view the Cali-
bration screen, which displays information about the most recent calibration
run. From this screen you can also configure Auto-Calibration parameters.
Figure 3-6.
Calibration screen.
Name
Conc.
Area
Rfactor
RT
Conc. Factor
3-8 | ta7000R Gas Purity Monitor
Idle Mode
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
This is the gas species to be measured.
This is the cylinder concentration (or blend) of the Calibration gas con-
nected to the instrument.
This is the area of the last analysis run for each gas type.
This is the Response Factor created by the most recent (or previous) calibra-
tion cycle. Rfactor is used to generate the area.
This is the Retention Time, which is the time point of the peak from the
start of the most recent analysis run.
This is a value that is preset at the factory. This number represents the
internal gas blender ratio.
Calibration
Name
Conc.
Area
8100
H2
*
7500
CO
Conc. Factor 200
Cal Time
00:00
Cal Purge
1 Min
Edit
12:34:25
Rfactor
RT
1346165
11737367
Interval
None
Cal Tol
10%
Page
PN T900-013, Rev A