Carrier Gas Supplies; Gas Ports; Sample Gas; Calibration - Ametek ta3000R User Manual

Gas purity monitor
Table of Contents

Advertisement

Carrier Gas Supplies (On-Line Installation)

Gas Ports

Sample Gas

Calibration Gas
Chassis
14 | Model ta3000R Gas Purity Monitor
Inlet Pressure Range
Inlet Pressure Stability
Return Pressure
Flow Rate
Temperature
Maximum Impurity Levels
Sample Inlet
Carrier
Sample Vent
Aux
Inlet Fitting
Flow Rate
Inlet pressure stability
Vent pressure
Inlet Fitting
Cylinder concentration
Cylinder volume
Dimensions
Weight
Power
70 to 90 psig (4.8 to 6.1 bar)
± 2%, regulator required
Atmospheric vent is optimal, ± 0.5 psig
maximum
50 cc/min minimum, bypass at 50
cc/min.
60 to 100 °F (16 to 38 °C), optimum when
maintained ± 2°C
Varries by application. An external purifier
may be required
1/16-inch VICI compression fitting
1/16-inch VICI compression fitting
1/16-inch VICI compression fitting
1/16-inch VICI compression fitting
1/16-inch VICI compression fitting
20 - 100 sccm minimum
± 2%, UHP regulator required
Atmospheric pressure vent is optimal, ± 0.5
psig maximum
1/16-inch VICI compression fitting
(Uses Sample Inlet)
H
, CO, Balance nitrogen, depends
2
on level of detection
100 cubic feet minimum (2800L)
7" (18cm) H x 16.8" (43cm) W x 26.5"
(67cm) D
35 lb. (15.9 kg)
100 - 120 VAC, 50/60 Hz; 200-240 VAC,
50/60 Hz