Error Condition / Description / Corrective Action

Event Program Error

Internal event program has not been correctly entered, is missing, or has been corrupted.
Corrective Action:
Contact AMETEK Service or qualified service representative for troubleshooting assistance.

Printer Error

Printer is not connected, is turned Off, or is not functioning properly.
Corrective Action:
• Check to ensure a printer is connected properly at the printer and the COM port on the
• Check the printer to ensure it has been powered On.
• Check the printer cable for damage.
• Test the printer to ensure it is working properly.
• Check to ensure the printer has paper.

Database Change, Update NV RAM

This fault will occur if any of the system parameters have been changed. The most common
system parameters that are changed are the response factor due to performing a calibration
run. This fault can only be updated during the Warm-up or Idle modes.
Corrective Action:
• Place the ta7000R into Warm-up mode to check if this updates and clears this error.
• Place the ta7000R into Idle mode to check if this updates and clears this error.
• If changes to the System Settings have been made, make sure they have been saved. Refer
PN T900-013, Rev A
ta7000R Rear Panel.
to "Saving Changes to Non-Volatile RAM" in Chapter 3 for information about how to perma-
nently save the changes.
|
Troubleshooting
4-3