Error Messages

Error Messages are viewed and acknowledged on the System Settings screen. When
the Status LED on the front panel is red this indicates that there are system errors or the
status of the analyzer has changed. The System Status page displays error messages that
the system has encountered. See the Run section for more information on the System
Settings page (35).
Chassis Controller Error
Internal electronics failure - Call manufacturer or qualified service repre-
sentative.
Detector Controller Error
Internal electronics failure - Call manufacturer or qualified service repre-
sentative.
Temperature Out of Range
Either the temperature zone is 2 degrees over the set point, or the tem-
perature zone has been disabled. Perform the Warmup procedure to clear
this error.
Vsig Out of Range (Lamp Off?)
During the Warmup cycle the analyzer performs a system test of the RGD
lamp. If the RGD lamp is not within operating range this error will be
displayed
Carrier Pressure Out of Range
External carrier pressure is not within specified limits. Verify external car-
rier pressure supply.
Analog Signal Zeroing Error
The instrument has failed to zero the detectors analog signal.