Emerson Rosemount 3410 Series Maintenance And Troubleshooting Manual page 21

Gas ultrasonic flow meters
Hide thumbs Also See for Rosemount 3410 Series:
Table of Contents

Advertisement

Maintenance and Troubleshooting manual
00809-0200-3104
Table 2-2: Troubleshooting
Error
Configuration changed
Configuration lost
Electronics Temperature
is Out of Nominal Range
Rosemount 3415, 3416 and 3417
(continued)
Recommended action(s)
One or more parameters have been modified in the meter's
configuration,
Collect an Audit log using MeterLink in order to see what
configuration parameters changed and when they changed.
Run the Tools → Edit/Compare Configuration utility and select
Write All or Write Checked values to write the changes to the
meter.
Save the configuration file.
The meter configuration has reset to default values and the meter is
not configured correctly to measure flow and the meter has
performed a Cold Start.
Unless the Cold Start occurred after upgrading firmware,
replace the CPU board.
If the cold start occurred after a firmware upgrade, fully re-
configure the meter from a previously saved configuration
using the Tools → Edit/Compare Configuration in MeterLink.
Temperature of the internal electronics is out of nominal operating
range (below -40 °C or above 100 °C) which could lead to a system
failure.
Attempt to warm or cool the meter electronics housing.
If the electronics is mounted to the meter and the process fluid
in the meter is over 60 °C, you must remote mount the
electronics off of the meter body.
Collect a Maintenance log using MeterLink while the meter is
experiencing the issue and contact your service representative.
Troubleshooting
April 2022
21