Table 2-2: Troubleshooting
Error
Configuration changed
Configuration lost
Electronics Temperature
is Out of Nominal Range
Rosemount 3415, 3416 and 3417
(continued)
Recommended action(s)
•
One or more parameters have been modified in the meter's
configuration,
—
Collect an Audit log using MeterLink in order to see what
configuration parameters changed and when they changed.
—
Run the Tools → Edit/Compare Configuration utility and select
Write All or Write Checked values to write the changes to the
meter.
—
Save the configuration file.
•
The meter configuration has reset to default values and the meter is
not configured correctly to measure flow and the meter has
performed a Cold Start.
—
Unless the Cold Start occurred after upgrading firmware,
replace the CPU board.
—
If the cold start occurred after a firmware upgrade, fully re-
configure the meter from a previously saved configuration
using the Tools → Edit/Compare Configuration in MeterLink.
•
Temperature of the internal electronics is out of nominal operating
range (below -40 °C or above 100 °C) which could lead to a system
failure.
—
Attempt to warm or cool the meter electronics housing.
—
If the electronics is mounted to the meter and the process fluid
in the meter is over 60 °C, you must remote mount the
electronics off of the meter body.
—
Collect a Maintenance log using MeterLink while the meter is
experiencing the issue and contact your service representative.