Maintenance and Troubleshooting manual
00809-0200-3104
6. Wait for 30 - 60 seconds to let the coupling fluid spread more evenly on the surface.
7. Insert the transducer capsule assembly into the transducer stalk while aligning the
8. Make sure that the indicator ring on the transducer capsule assembly is flush with
9. Screw the transducer retainer on the stalk and use a 1 ⅛" wrench to bottom out.
10. Connect the transducer chordset to the transducer retainer. The internal connector
11. Secure the transducer cabling nut by turning clock wise. Ensure the cable nut
12. Repeat
13. Continue with
Rosemount 3415, 3416 and 3417
Note
Do not press the Kapton tape while spreading the coupling fluid.
anti-rotation pin on the capsule with the slot on the inner surface of the stalk.
Figure 3-13: Length verification of T-200 transducer capsule assembly
A. Transducer stalk
B. Indicator ring
C. Capsule assembly
the end of the transducer stalk (see
correct length.
Note
Do not torque down the transducer retainer to avoid damages to the anti-rotation
pins.
of the transducer chordset is keyed and will only go on one way.
threads are correctly aligned.
Step 1
through
Step 11
capsule assemblies were replaced.
Modifying the calibration parameters for T-Slot transducers
the MeterLink Transducer Swap-out Wizard to update delta delay time.
Figure
3-13) to ensure the capsule is set to the
for other transducer assemblies if their transducer
Meter repairs
April 2022
to use
65