Brief Sensor Issue Alert; Technical Alerts - Dexcom G7 User Manual

Hide thumbs Also See for G7:
Table of Contents

Advertisement

App
No Alerts
Brief Sensor Issue
Do not remove sensor.
Temporary issue.
Wait up to 3 hours.
Help
Sensor Failed
No Alerts
Remove Sensor Now
To remove sensor, peel
off patch from the edge.
Technical alerts can't be turned off, but you can change how your alert notifies you:
App: For each alert, you can choose between Sound, Vibrate, Match Phone
l
Settings.
Vibrate Exceptions: Urgent Low and technical alerts act differently; if you don't
acknowledge them, they will add sound. In the app, these alerts include: Urgent
Low, App Bluetooth is Off, App is Closed, App Location is Off, App Stopped: Phone
Storage Full, App Stopped Working, Phone Bluetooth is Off, Phone Location is Off,
Replace Sensor Now, Sensor Failed
If you use a Quiet Mode (Silence All or Vibrate), it affects all alerts, including

technical Alerts.

Silence All Exceptions: App Stopped Working and App Stopped: Phone Storage
Full alerts will still sound.
Receiver: Choose between Silence All and Vibrate Only
l
Vibrate Only Exception: Urgent Low and technical alerts act differently; if you
don't acknowledge them, they will add sound. In the receiver, these alerts include:
Urgent Low, Replace Sensor Now, Sensor Failed, Set Date/Time, System Check,
Very Low Battery, Weak Charger.
5 • Alerts |
Dexcom G7 User Guide
Receiver
Brief Sensor Issue
You won't get sensor
readings or alerts.
Please wait...
Sensor Failed
Replace your sensor now.
No sensor readings or alerts
until you replace sensor.
What it means

Brief Sensor Issue alert

Alerts you when the sensor can't read your
glucose right now.
Technical alerts
Alerts you when you're not getting sensor
readings. They include Sensor Failed,
Replace Sensor Now, and similar alerts.
49

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the G7 and is the answer not in the manual?

Questions and answers

Melanie Binns
February 26, 2025

Where can you place the sensor

1 comments:
Mr. Anderson
February 26, 2025

You can place the Dexcom G7 sensor on:
- The back of the upper arm
- The abdomen (not for pregnant women)
- The upper buttocks (for ages 2-6 years only)

Ensure the site is at least 3 inches (8 cm) from any insulin pump infusion set or injection site, away from waistbands, scarring, tattoos, irritation, and bones, and in a spot unlikely to be bumped, pushed, or laid on while sleeping.