When HIGHER is better
Adults
93.6%
(>99.9%)
*Reference is YSI (Yellow Springs Laboratory Instrument)
Insertion comfort
Participants reported on a questionnaire that 95% of sensor insertions felt painless
(mild, no pain).
Product specifications
Use electrical equipment as directed:
Use of accessories, cables, adapters, and chargers other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
Portable radio frequency communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 centimeters
to any part of the Dexcom G7 CGM System including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
124
Performance Metrics*
Clinical Accuracy
% of readings in Consensus Error Grid
Zone A
(%CEG Zone A+B)
Glucose readings in Zone A are
expected to result in treatment
decisions considered clinically accurate,
while readings in Zone B are thought to
pose minimal risk.
Pediatrics
93.2%
(99.8%)