1.J. Environmental Requirements (Operating, Shipping and Storage Conditions)
Power Requirements:
Temperature:
Humidity:
Operating Altitude:
The Maxwell
®
shortening the expected lifespan of the instrument, install in a location that meets the following criteria:
Locate on a sturdy, level surface.
Avoid dusty areas.
Choose a location that has good air circulation and is not exposed to direct sunlight.
Avoid noisy electrical power sources (e.g., power generators).
Do not install in a location where there is large temperature variability or high humidity.
Do not position the instrument so that it is diffi cult to unplug from the power source.
Do not place next to heat sources.
Do not use near fl ammable gases or liquids.
Do not place near other electrically sensitive instruments.
Connect the Maxwell
to ensure continued operation during brief power outages that otherwise might result in aborted
instrument runs and lost samples.
2.

Hardware Overview

Figure 2. Front of Maxwell
10
Promega Corpora on · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM424 · Revised 2/15
95–240VAC, 50/60Hz, 1.0A
4°C–50°C (shipping/storage), 15°C–25°C (operation)
Up to 80% relative humidity, non-condensing
<2000 meters
CSC Instrument is intended for indoor use only. Wipe up spills immediately. To avoid
®
CSC Instrument and Tablet PC to an uninterrupted power supply (UPS). This is
®
CSC Instrument.