User Aborts Protocol
If the protocol is aborted by the user, the 'Protocol Running' window will indicate that the protocol has been
aborted (Figure 50). After a protocol is aborted, the Current Step will be listed as Aborted by user. After Aborting
the protocol, you can press the Open Door button to bring up the 'Clean Up' window (Figure 51).
Figure 50. Protocol Running window after protocol abort. If a protocol is aborted by the user or due
to an instrument error, the 'Protocol Running' window changes to indicate that the protocol has been
aborted and specifi es the reason next to the Current Step: title. After a protocol is aborted you can touch the
Open Door button to run the Clean Up procedure. An aborted run (initiated by the user or due to instru-
ment error) will result in the loss of all the samples. Do not attempt to re-purify samples from an aborted
run.
Instrument Error
If the protocol is aborted due to instrument error, the 'Protocol Running' window will change to indicate
that the protocol has been aborted (Figure 50) and will display an error message.
After a protocol is aborted the Current Step will list the reason for aborting the protocol.