5.C. Clean Up
After Aborting the protocol, you can press the Open Door button to bring up the 'Clean Up' window
(Figure 51).
The 'Clean Up' window asks you to check to see if plungers are still engaged on the plunger bar. If the
plungers are not engaged, remove the deck tray from the instrument and touch the Skip Clean Up button to
continue. On pressing the Skip Clean Up button, you will be presented with the extraction report
(Figure 52).
Figure 51. Clean Up window after protocol abort/instrument error. If a protocol is aborted by the
user or due to an instrument error, the 'Clean Up' window is displayed that asks the users to Start or Skip
Clean Up depending on whether the plungers are still engaged or disengaged from the plunger bar.
If some or all of the Plungers are still engaged on the plunger bar, you must perform the following steps to
remove the plungers before another purifi cation run can be performed:
Remove cartridges containing ejected plungers from the deck tray.
Reinsert the deck tray with the remaining cartridges (those missing ejected plungers).
Touch the Start Clean Up button to eject the remaining plungers .
After the Clean Up is successful, you can press the Open Door button and remove the deck tray. You will be
presented with the 'Report View' window.
If the plunger cleanup fails, you should contact Promega Technical Services for further assistance.