1.E. Environmental Requirements
Operations, Shipping and Storage Conditions
Operating Conditions
Operating Altitude
Shipping and Storage Conditions
Maximum Relative Humidity
MAINS Power Supply
Transient Overvoltage
Temporary Overvoltage
Pollution Degree
The instrument is intended for indoor use only.
Install on a sturdy, dust-free, level surface.
Do not install in a location where there is large temperature variability or
high humidity.
Position the instrument so that it is easy to unplug from the power
source.
Do not place the power supply in areas where it may come in contact
with liquid.
Do not place next to heat sources.
Do not use near flammable gases or liquids.
For optimal performance, we recommend equilibrating the instrument to
operating temperature before use.
2.
Quantus™ Fluorometer Inspection and Setup
Upon receiving your Quantus™ Fluorometer, inspect everything carefully and
make sure all items are present and undamaged.
2.A. Setting Up the Quantus™ Fluorometer
1.
Position the Quantus™ Fluorometer on a flat surface.
2.
Select the appropriate outlet adapter for your wall outlet from the Outlet
Adapter Kit provided.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Part# TM396
Page 6
15–30°C, up to 75% humidity
Up to 2,000 meters
5–40°C under noncondensing
conditions, up to 75% humidity
Up to 80% for temperatures up to
31°C, decreasing linearly up to 50%
relative humidity at 40°C
Supply voltage fluctuation up to ±10%
of the nominal voltage
Up to the level of Overvoltage
Category II
Occurring on the MAINS power
supply
Pollution Degree 2
Printed in USA.
Revised 1/20