 Clean filter chamber.
 It is now also possible to rod through the probe tube in order to remove deposits. Caution danger of breakage
when handling probe tubes made of alu-oxide !
 Replace filter holder with lid, turn handle C into position F and tighten lid with toggle A again.
 Replace the protective hood and fasten with quick-release fasteners.
 Open (if available) ball valve.
17.2

CHANGE OF THE OPTIONAL PREFILTER

For changing the prefilter the complete probe unit has to be dismounted and removed out of the process:
Warnung
Loosen the 4 flange nuts and remove if possible the fixing screws (not for prefilters V12..., adapter flanges
or SP30-H....)
Extract probe with prefilter out of the process.
Let prefilters cool and than screw out resp. for prefilters V12... remove the 4 flange nuts on probe side (see
also Fig. 6).
Change or clean prefilter.
Note
www.mc-techgroup.com
When working during operation:
High surface temperatures!
Touching the surfaces can result in burns. Wear protective gloves and any unau-
thorized access to the probe must be made impossible!
Prefilters depending on type and degree of contamination can be partly cleaned
mechanically or in an ultra sonic bath and reused.