Figure 3
Cross-sectional drawing SP2000-H
Removing the filter housing lid to check or change the filter element has to be done as follows :
 Remove the probe's protection shield after opening the two clamping devices;
 Turn toggle screw „A" approx. 1 turn counter-clockwise, that the filter housing lid will lift up;
 Turn handle „C" in position „E";
 Turn clamp „B" counter-clockwise in direction „G";
 Take toggle screw „A" and pull out the filter housing lid;
 Check tight fit of filter element and tighten it if necessary (hand tight) or change filter element and correspond-
ing sealings;
Warning
 Push in the filter housing lid into the filter area;
 Turn clamp „B" clockwise and turn with the handle „C" the ringsscrew „D" in position „E", that the clamp „B" will
latch into the ringscrew „D" and the threaded bolt „H". It could be neccessary to move the filter housing lid a
little bit forward and backward;
 Turn handle „C" in position „F" and fasten the filter housing lid by turning the toggle screw „A" clockwise by
hand.
www.mc-techgroup.com
When working during operation:
High surface temperatures!
Touching the surfaces can result in burns.
Wear protective gloves and any unauthorized access to the probe must be made
impossible !
SP2000, SP2300, SP2400 | 1.00.03