RECEIPT OF GOODS AND STORAGE

 The gas sample probe and any special accessories should be removed carefully from the packaging and checked
immediately for completeness against the delivery note.
 Check the goods for any damage incurred during transport and if necessary inform your transport insurer of any
damage.
The gas sample probe is normally delivered in two packaging units:
1. The gas sample probe with the screws, nuts and flange seal required for mounting.
2. Sample tube or prefilter, if applicable with extension tube.
The equipment should be stored in a protected, frost-free room!
Note
13

PREPARATION FOR INSTALLATION

 Select the optimal sampling point in accordance with the generally applicable guidelines or consult the com-
petent persons.
 Locate the sampling point in such a way that there is adequate space for inserting and removing the probe and
pay attention to the insertion length of the probe tube.
 Make certain that the probe is easily accessible so that you can carry out any subsequent maintenance work
without trouble.
 Locate the probe connections in such a way that the connections' temperature is always above the acid dew
point in order to avoid corrosion and blockage problems. If this is not possible, a heated SP35/SP30 probe tube
is recommended for cold connections.
 If the ambient temperature in the area of the connections is > 80 °C [176 °F] as a result of radiated heat, then a
radiated-heat deflector must be mounted to protect the probe.
 The connection's mounting flange connection should comply with DN65 PN6. If other connection sizes are
required, a special adapter flange /S010 can be supplied as an option. Instead of flange connection mounting,
the probe can also be mounted using a /SO1 R2" adapter on a corresponding threaded sleeve connection. This
adapter can be supplied. The necessary minimum flange size and the minimum connection diameter depends
on the diameter of the probe tube or prefilter used.