LABNET SPECTRAFUGE 6C

3 .0 Specifications

Maximum Speed
Maximum RCF
Maximum Volume
Admiss. Density
Electrical / Fuse Rating
Dimensions (W x D x H)
Operating Conditions

4.0 Package Contents

Description
Spectrafuge™ 6C
Angle rotor for 6 x 15 mL tubes
Rotor wrench, 7/16
Power Cord

5.0 Installation

The centrifuge should be installed on a rigid, even surface such as a stable laboratory bench, cabinet,
etc. To guarantee sufficient ventilation, ensure that the centrifuge has at least 6 inches (15 cm) of free
space on all sides, including the rear.
The centrifuge should not be located in areas subject to excessive heat such as in direct sunlight or
near radiators or the exhaust of a compressor, as a buildup of heat may occur within the chamber.
Before operating the centrifuge, check that the power source corresponds to that on the
manufacturer's rating label, then connect the power cord to the centrifuge and the power source.
6,500 rpm
4,000 x g
6 x 15 mL
1.2 kg/dm
3
120V, 50-60Hz, 1.0A/1.25AT
230V, 50-60Hz, 0.6A/630mAT
8.25 x 9.5 x 7.0 in. (21 x 24 x 18 cm)
5°C to 40°C / ≤80%RH
Quantity
1 each
1 each
1 each
1 each