Dimensions (mm)
17 x 120
Capacity*
15 mL
Tube image
Radius (cm)
8.2
Max. RFC
3,873
Tubes per Rotor
6
Adapter (C0200-17A)
–
*Capacity may vary based on serum volume
**Sarstedt is registered trademark of Sarstedt Company. The compatibility of these tubes has been independently determined by Corning
and there should be no implication of sponsorship or endorsement by Sarstedt Company.
6.0 Installation of Rotors and Rotor Maintenance
The following accessories are available for the Spectrafuge™ 6C:
Angle rotor for 6 x 15 mL tubes
Cat. No.
Tube Measurement
Maximum Speed
Centrifuging Radius
RCF
Adapter for 5 mL (12 x 75 mm) and 7 mL (13 x 100 mm) tubes
Cat. No.
Tube Measurement
Tube and tube adapter reference chart:
Standard Blood
Collection Tubes
16 x 100
13 x 100
12 x 75
10 mL
7 mL
5 mL
8.5
7.7
7.7
4,015
3,637
3,637
6
6
6
–
Yes
Yes
6.1 Rotor Maintenance
The rotor should be cleaned thoroughly after each use. Thorough cleaning must be performed when
spinning samples containing phenol or phenol chloroform. Periodically inspect the rotor for dents,
dings, scratches, discoloration and cracks If any damage to the rotor is found, discontinue use of the
rotor immediately and replace.
6.2 Removing and Installing the Angle Rotor
Remove the rotor screw from the motor shaft by turning the screw counterclockwise. Lift the rotor
upward and remove from the centrifuge.
Ensure that the motor shaft adapter remains on the motor shaft (Figure 1). Clean the motor shaft and
motor shaft adapter (see figure 1).
Place the rotor on the motor shaft (figure 1) and over the motor shaft adapter (see figure 1 and 2).
NOTE: Figure 1 and 2 are located on the following page.
When loading the rotor, refer to figure 3 (located on the following page). Loading in the pattern
indicated will ensure a balanced load. Tubes to be loaded should be filled equally by eye. The difference
in the weight between the tubes should not exceed 2-3 grams. A partially loaded rotor may be
centrifuged if the loading scheme for balancing a rotor given in figure 3 is followed.
TR100
15 mL (17 x 120 mm) - 10 mL (16 x 100 mm)
6,500 rpm
8.5 cm
4,000 x g
C0200-17A
12 x 75 mm, 13 x 100 mm, or common Sarstedt™** style tubes
15 x 102
16 x 92
15 x 92
10 mL
9-10 mL
7-8 mL
8.5
8.5
8.5
4,015
4,015
4,015
6
6
6
–
–
–
Sarstedt** Blood
Collection Tubes
13 x 90
11 x 92
11 x 66
4.9 mL
4-5 mL
2-3 mL
7.7
7.7
7.7
3,637
3,637
3,637
6
6
6
Yes
Yes
Yes
Instruction Manual
13 x 65
8 x 66
2-3 mL
1-1.4 mL
7.7
7.7
3,637
3,637
6
6
Yes
Yes
|
3
Need help?
Do you have a question about the Spectrafuge 6C and is the answer not in the manual?
Questions and answers