GE Healthcare Brivo XR118 Service Manual page 301

Table of Contents

Advertisement

GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
Change the DI Display
1. On the Deviation Index screen, click the DI Display drop-down list to open
it.
2. Select the display option.
 Available options are:
3. Press [Save] to apply the change and close the screen.
 [Close] closes the screen without saving the changes.
Change the Target EI (Exposure Index)
1. On the Deviation Index screen, press an anatomical category to expand it.
2. Press the category again to close it.
3. Change the Target EI as appropriate for the view.
4. Repeat for all applicable anatomical categories.
5. Click [Save] when finished to apply the changes and close the screen.
 [Cancel] closes the screen without saving the changes.
6
EVISION
– N
o
Display
– Numerical
– Graphical
Chapter 12 - - Set Preferences
B
XR118 S
RIVO
In this example, there is no DEI
information displayed on the Image
Viewer screen.
In this example, only numerical DEI
information displayed on the Image
Viewer screen.
In this example, DEI information is
displayed both numerically and
graphically on the Image Viewer screen.
This is the recommended display
option.
M
ERVICE
ANUAL
Page 301

Hide quick links:

Advertisement

Table of Contents
loading