Chapter 4 - Target System Check
Digital X-Ray Imaging System (Brivo XR118) can upgrade target analog system which meets the
below requirements to DR system.
Before installation XR118, purchaser needs to check target system from two sides.
1.) In order to reduce installation time, please call customer to check target system before go to
2.) Make sure the target system meets all the requirements below. For more information, please
Table 4-1 Target System Requirements
No.
1
2
3
4
5
6
7
8
9
6
EVISION
site, including:
a.) The system can be installed into Wall stand/Table. Make sure the cassette on target
system can hold the detector (Width: 383.5±1mm, Height:459.5±1mm).
b.) Target system can make exposure.
c.)
Target system is fix RAD system.
d.) Target system's Generator frequency is higher than 400Hz.
e.) Suggest the hospital to ask RIS engineer also to be on site for DICOM connectivity in case
RIS query issue occurs, it often needs RIS side's update to meet DICOM3.0.
contact your sales representative.
Categor
Item
y
Generat
Maximum kV
or &
Maximum mA
Tube
Maximum mAs
Maximum exposure
time
Generator inverter
switching frequency
Generator
Tube
Cassette
Size
Tray
Grid
Line Pair
Chapter 4 - - Target System Check
B
RIVO
Preferred Capability
>=80kV, kV accuracy <=10%
>=50mA, mA accuracy <=20%
>3.2mAs, mAs accuracy <= ±
(10% + 0.2mAs)
<=10s, exposure time accuracy
<=± (10% + 1 ms)
>=400Hz
Comply with GB9706.3-2000/
IEC60601-2-7:1998 or IEC
60601-2-54: 2009
Comply with GB9706.11-1997/
IEC601-2-28:1993 or IEC
60601-2-28: 2010
Compatible with ISO4090
35cmx43cm, Thickness 16mm
cassette
36, 40, 60, 70, 78lp/cm
XR118 S
M
ERVICE
ANUAL
Target
Pass
System
/Fail
Claim
Page 43