Removing The Etco2 And Nibp Modules - ZOLL Propaq M Service Manual

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

Removing the EtCO
Tool Required
# 1 Phillips screwdriver with torque capability
orange (wooden) stick
Setup
This procedure should be conducted on a static mat.
Wear a wrist strap to ground during the procedure.
Complete these procedures in the following order before performing this procedure:
1.
2.
3.
Note:
To assemble the EtCO
EtCO
and/or NIBP Disassembly
2
1
2
9650-0804-01 Rev. A
and NIBP Modules
2
"Removing the Handle" on page 43
"Disassembling the Main Unit" on page 44
"Removing the Front Enclosure Module" on page 47
After replacing the EtCO2 and/or NIBP modules, see Chapter 4, "Resetting Life Cycle
Counters" to reset the life cycle counters for the module(s) that you have replaced.
and/or NIBP modules, reverse the steps.
2
Remove the right side panel (containing
temperature, IBP, and USB connectors) by
removing four (4) screws from inside the
unit.
Note: During reassembly, torque screws to
6 in-lbs.
Remove three (3) screws from underneath
and outside the unit.
Note: During reassembly, torque screws to
6 in-lbs.