Error Message
NIBP AIR LEAK
NIBP EQUIPMENT CHECK
NIBP WEAK PULSE
NIBP TEMP OUT OF
RANGE
NIBP ARTIFACT
9650-0804-01 Rev. A
Explanation
Major air leak or could not
reach target pressure.
Device is unable to obtain a
reading.
Too few pulses are detected
during the reading attempt.
Module temperature is out of
its specified operating range.
Unable to determine blood
pressure. Check
connections; restrict patient
movement.
ZOLL Propaq M Service Manual
Propaq M Error Messages
Technical Action
Attempt to identify
problems with the hose
and/or cuff by doing the
following:
Check the hose
connection to the
device. Make sure that
the hose is tightly
screwed onto the NIBP
connector.
Ensure that the cuff is
appropriately wrapped
and is not too loose.
Inspect the cuff and
hose for leaks. Inspect
the O-ring on the NIBP
hose connector.
Consider replacing the
cuff and/or hose.
If the problem persists,
perform one of the following
procedures in Chapter 1,
"NIBP Valve, Pump, and
Tubing Integrity Tests":
NIBP Leak Test
NIBP Inflation Test
Inspect internal NIBP
tubing for leaks.
Check the hose and cuff
connections. If problem
persists, perform one or
more of the following
procedures outlined in
Chapter 1, "NIBP Valve,
Pump, and Tubing Integrity
Tests":
NIBP Leak Test
NIBP Deflation Test
NIBP Inflation Test
Tighten the cuff and ensure
appropriate placement.
Move the device to a cooler
or warmer environment.
Attempt to identify and
eliminate the source of
artifact. If problem persists,
replace the NIBP module.