ZOLL Propaq M Service Manual page 34

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE
NIBP Test
NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinder
Tools Needed
Connect the Propaq M to the NIBP simulator using the single or dual lumen hose. Wrap the
Test Setup
NIBP cuff around a rigid cylinder, such as a glass jar.
Note:
Do This...
1.
On the Propaq M, ensure the patient mode
is set to Adult.
2.
Set the simulator to the 80/50 simulator
preset.
3.
Take a manual reading.
4.
Set the simulator to the 120/80 simulator
preset.
5.
Take a manual reading.
6.
Set the simulator to the 200/150 simulator
preset.
7.
Take a manual reading.
30
T
ESTS
Refer to Appendix B for information on simulators and settings that you may use to
assess the performance of the NIBP module.
Observe This...
Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
9650-0804-01 Rev. A
Pass/Fail
o o
o o
o o