C
1 M
HAPTER
AINTENANCE
Acceptable Results
The unit passes the test if P
or equal to 4.0 mmHg for all three target pressures.
Corrective Action
If readings fail to meet the required accuracy, the NIBP module requires replacement or factory
analysis and realignment.
8
T
ESTS
is less than or equal to 3.0 mmHg and P
d-Primary
is less than
d-Safety
9650-0804-01 Rev. A