ZOLL Propaq M Service Manual page 16

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE
Corrective Action
If the unit fails the test, the most likely cause is damaged or disconnected internal tubing. The
device needs to be disassembled to replace or reattach the appropriate segment of NIBP tubing.
The NIBP valves should also be inspected.
12
T
ESTS
9650-0804-01 Rev. A