ZOLL Propaq M Service Manual page 33

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

IBP Test
IBP patient simulator (with accuracy better than +/- 1 mmHg), or equivalent
Tools Needed
Connect the test cables to IBP patient simulator.
Setup
Do This...
1.
Connect the IBP patient simulator to the
Propaq P1 channel.
2.
Set the IBP patient simulator to 0 mmHg.
3.
On the Propaq M, press > IBP.
4.
On the Propaq M, Press > P1->0<-.
5.
Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
6.
Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.
9650-0804-01 Rev. A
Observe This...
Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 1 is
present.
Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).
ZOLL Propaq M Service Manual
Functional Verification Tests
Pass/Fail
o o
o o
o o
29