ZOLL Propaq M Service Manual page 56

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

3
4
5
6
7
8
52
Disconnect multi-wire IBP cable with ferrite
in the area illustrated.
Cut the tie wrap to the IBP cable in the area
shown with the orange stick.
Note: During reassembly, secure a new tie
wrap to the IBP cable.
Disconnect the USB flex cable from ZIF
connector in the area illustrated.
Lay front end of separated unit down with
opened side facing up.
Disconnect the black EtCO
connector from the air vent.
Remove four (4) screws in the areas
illustrated to release the silver EtCO
module.
Note: During reassembly, torque screws to
6 in-lbs.
multi-wire IBP cable
tube
2
2
9650-0804-01 Rev. A