Error Message
CO2 GAS INPUT
OCCLUSION
CO2 MAINTENANCE
REMINDER
CO2 MAINTENANCE
REQUIRED
CO2 NO FILTERLINE
CO2 CAL FAILURE
9650-0804-01 Rev. A
Explanation
CO
inlet blocked.
2
Periodic maintenance
reminder.
Periodic Maintenance due.
No filterline is detected.
Calibration failed.
ZOLL Propaq M Service Manual
Propaq M Error Messages
Technical Action
®
Replace FilterLine
. If
problem persists, check for
CO
inlet occlusion. If still
2
persists, replace the CO
module.
Replace CO
module soon.
2
Replace CO
module.
2
If FilterLine is connected
and problem persists,
replace FilterLine. If still
persists, replace the CO
module.
Retry calibration or
continue with old
calibration. Replace the
CO
module.
2
2
2
39