ZOLL Propaq M Service Manual page 42

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

C
2 T
HAPTER
ROUBLESHOOTING
Error Message
NIBP NEONATAL
DETECTED
NIBP LOW VOLTAGE
SPO2 COM ERROR CODE
SPO2 SENSOR FAILURE
SPO2 UNKNOWN
RECOVERY NUM
CO2 DEVICE FAILURE
CO2 COM ERROR CODE
CO2 BAROMETRIC OUT
OF RANGE
CO2 CAL REMINDER
CO2 CAL REQUIRED
CO2 CHECK FLOW
38
Explanation
Device thinks a neonate cuff
is connected. If no neonate
cuff is connected, tubing
may be damaged or kinked.
Insufficient voltage is
supplied to the NIBP
module.
Communication error.
Defective sensor or sensor
error.
Recovery error.
Malfunction or device failure.
Communication failure.
Altitude is too high for
module.
Calibration reminder.
Calibration due.
Exhaust port blocked.
Technical Action
Verify that neonate cuff is
not connected. If problem
persists, check the routing
of the NIBP tubing inside
the device.
Check the module voltage
in the NIBP Test screen.
Inspect the unit or water
ingress. If evidence of
water ingress is found,
allow the water to
evaporate while periodically
checking the module
voltage. If problem persists,
replace the NIBP module.
Check communication
cables. If problem persists,
replace the SpO
module.
2
Replace the SpO
sensor.
2
Replace the SpO
sensor.
2
Replace the CO
module.
2
Check communication
cables. If problem persists,
replace the CO
module.
2
Bring device into specified
altitude range. If problem
persists, replace the CO
module.
Calibrate the CO
module
2
soon.
Calibrate the CO
module.
2
Check for exhaust port
occlusion. If problem
persists, replace the CO
module.
9650-0804-01 Rev. A