Chapter 8 Cleaning/Disinfecting/Storing

Cardiac Output Relay Cable

 Do not reuse / resterilize the cardiac output catheter.
 When cleaning, follow the procedure below.
1
Wipe the cable using 70% isopropyl alcohol cotton.
2
Dry it completely with air before reusing.

Airway Adapter for Capnostat 5

 Wash in lukewarm soapy water. Then dip it in antiseptic solution (ex. glutaraldehyde) for low-temperature
sterilization. Dry after rinsing in sterile water.
 Use EOG (Ethylene Oxide Gas) to sterilize. Proper ventilation must be performed.
 Before re-using the airway adapter, make sure the window is dry and no residue is left.
Check if the adapter is not damaged by the operation or cleaning / sterilization.
CAUTION
 Do not sterilize the airway adapter using autoclave methods.
 Do not reuse / re-sterilize disposable airway adapter.

BISx

 Clean any spillage of blood or solutions on BISx as soon as possible.
 Use lint-free absorbent towels for spill cleanups.
 Dampen the towel with detergent and lukewarm water to aid in cleaning.
 After cleaning, wipe the PIC connector ends with alcohol and allow to dry completely.
 Residual moisture inside the connector may affect BISx performance.
 Use lint-free absorbent towels dampened with a 10% bleach solution, or a commercial disinfectant.
CAUTION
 BIS sensor is disposable. Do not reuse it.
 Do not reuse the BIS sensor to other patients. It may cause cross-infection.
 The duration for one usage should be within 24 hours.
Cleaning the Equipment and Sensors
8-5