Chapter 8 Cleaning/Disinfecting/Storing
Cardiac Output Relay Cable
Do not reuse / resterilize the cardiac output catheter.
When cleaning, follow the procedure below.
1
Wipe the cable using 70% isopropyl alcohol cotton.
2
Dry it completely with air before reusing.
Airway Adapter for Capnostat 5
Wash in lukewarm soapy water. Then dip it in antiseptic solution (ex. glutaraldehyde) for low-temperature
sterilization. Dry after rinsing in sterile water.
Use EOG (Ethylene Oxide Gas) to sterilize. Proper ventilation must be performed.
Before re-using the airway adapter, make sure the window is dry and no residue is left.
Check if the adapter is not damaged by the operation or cleaning / sterilization.
CAUTION
Do not sterilize the airway adapter using autoclave methods.
Do not reuse / re-sterilize disposable airway adapter.
BISx
Clean any spillage of blood or solutions on BISx as soon as possible.
Use lint-free absorbent towels for spill cleanups.
Dampen the towel with detergent and lukewarm water to aid in cleaning.
After cleaning, wipe the PIC connector ends with alcohol and allow to dry completely.
Residual moisture inside the connector may affect BISx performance.
Use lint-free absorbent towels dampened with a 10% bleach solution, or a commercial disinfectant.
CAUTION
BIS sensor is disposable. Do not reuse it.
Do not reuse the BIS sensor to other patients. It may cause cross-infection.
The duration for one usage should be within 24 hours.