If the reading does not match the tile value check the value stored in the gauge memory. See "Changing
calibration value" on page 23.
7 SET-UP OPTIONS
7.1 CALIBRATION VALUE
The calibration value is the value that is stored in the gauge to calibrate in the automatic mode. This value
must match the value on the label of the gloss standard.
Normally the value does not need to be altered before use, as it will be set before the gauge is delivered.
If it becomes necessary to change the stored value, for example if the standard is re-calibrated or a different
standard is used, then use the following procedure.
CHANGING CALIBRATION VALUE
1. From the measurement mode, press the SCROLL UP/DOWN key until the display shows:
System Set Up
Press Read Key
2. Press the READ/SELECT key to enter this menu option. The display will show:
SET STD CAL VAL.
60° 93.2