3. Press the READ/SELECT key to activate the value setting option. The display will show:
4. Use the SCROLL UP/DOWN key to adjust the value until it matches the value quoted on the standard
tile that is in use.
5. When the displayed value is correct press the READ/SELECT key to store the value. The gauge will
confirm this with the following message:
The gauge will then return to the measurement mode and the gauge must be re-calibrated before it is used
for gloss measurements - see "Calibration" on page 15.
7.2 SWITCHING BACKLIGHT ON/OFF
To extend the life of the batteries it is possible to switch the backlight off. The backlight can be switched on
and off using the following procedure:
1. From the measurement mode, press the SCROLL UP/DOWN key until the display shows:
2. Press the READ/SELECT key to enter this menu option.
24
SET STD CAL VAL.
Mode Active
CAL DATA STORED
Completed
System Set Up
Press Read Key