  • Page 1 Elcometer 401 Novo-Gloss™ Glossmeter Operating Instructions...
  • Page 2 (in a retrieval system or otherwise) or translated into any language, in any form or by any means (electronic, mechanical, magnetic, optical, manual or otherwise) without the prior written permission of Elcometer Instruments Ltd. TMA 0252 Issue 02 Text with Cover No: 16708...
  • Page 3: Table Of Contents

    CONTENTS Section Page About your gauge ....... . . 2 1.1 Standards .
  • Page 4: About Your Gauge

    Different surfaces will have different levels of gloss and require different reflective angles. The user can calibrate the Elcometer 401 using a certified calibration tile supplied with the gauge. 1.1 Standards Your Elcometer 401 Novo-Gloss™...
  • Page 5: What This Package Contains

    Consult your local Environmental Authority for further guidance. Note: Before you use your Elcometer 401 for the first time, you must charge the battery for a minimum of 16 hours. To maximise the benefits of your new Elcometer 401 please take some time to read these Operating Instructions.
  • Page 6: Getting Started

    Measurement aperture Figure 1. Elcometer 401 Controls and Features 2.2 Charging the battery Plugging the charger into the socket marked DC on the side of the gauge allows the battery to be charged. The battery will be fully charged in approximately 16 hours from full discharge.
  • Page 7: Taking A Reading

    3 TAKING A READING 1. Inspect the underside of the gauge to be sure it is clean and free of scratches. Inspect the lenses and the channels leading to them and remove any dust that may accumulate in use, storage or transit.
  • Page 8: To Calibrate Your Gauge

    Regular cleaning of the surface of the tile using the cloth provided is recommended. If necessary the tile can be cleaned using warm soapy water and dried with a soft lint-free cloth. Foam Lid - Zero Gloss Value Calibration Tile - High Gloss Value Figure 2.
  • Page 9: Gloss Measurement

    Best results are obtained on cylinders when measuring parallel to the axis. For accurate gloss measurements on curved surfaces with a radius as low as 12 mm (½") a special gauge, the Elcometer 400 NOVO-CURVE, is available - see “Related equipment” on page 14.
  • Page 10: Gloss Measurement

    In the unlikely event of a fault, the Elcometer 401 should be returned to your local Elcometer supplier or direct to Elcometer. Contact details can be found on the outside cover of these instructions and at www.elcometer.com 6.1 Calibration tile...
  • Page 11: Gauge

    BAM (Germany) or NRC (Canada). 6.2 Gauge The Elcometer 401 should always be checked to ensure that the metal base is clean. Care must be taken to ensure that the base remains free of any damage. The optical cavities and lenses must also be free of dust or any other contamination.
  • Page 12 Single Angle Glossmeter Remove the gauge case by unscrewing the four fixing screws found at the back of the gauge. Withdraw the gauge from the case. Unscrew the two lamp clamping ring screws, disconnect the inline connector and remove the old lamp. Fit the new lamp in the orientation shown in Figure 3.
  • Page 13 Dual Angle Glossmeter Note: If changing more than one lamp, mark the inline connectors to ensure correct connection. Remove the gauge case by unscrewing the four fixing screws found at the back of the gauge. Withdraw the gauge from the case. Unscrew the two lamp clamping ring screws holding the faulty lamp.
  • Page 14 Triple Angle Glossmeter Note: If changing more than one lamp, mark the inline connectors to ensure correct connection. Remove the gauge case by unscrewing the four fixing screws found at the back of the gauge case. Withdraw the gauge from the case. Unscrew the two lamp clamping ring screws holding the faulty lamp.
  • Page 15: Storage And Transit

    Always store the gauge in its case when it is not being used and during transit. 8 SPARES The Elcometer 401 Novo-Gloss™ Glossmeter is complete with all the items required to get started and take measurements, however over the life of the gauge replacements may be required. The following replacement items are available from Elcometer or your local supplier.
  • Page 16: Technical Specification

    10 RELATED EQUIPMENT In addition to the Elcometer 401 Novo-Gloss™ Glossmeter, Elcometer produces a wide range of other equipment for measuring the appearance of surface coatings. Users of the Elcometer 401 may also benefit from the following Elcometer products: •...