Endress+Hauser Fieldbus Levelflex M FMP41C Operating Instructions Manual page 54

Run time measurement
Hide thumbs Also See for Fieldbus Levelflex M FMP41C:
Table of Contents

Advertisement

Commissioning
ENDRESS + HAUSER
+
E
ENDRESS + HAUSER
+
E
!
54
Function "dist./meas.value" (008)
The distance measured from the reference point to the product surface and the meas. value
calculated with the aid of the empty alignment are displayed again. Check whether the values
correspond to the actual meas. value or the actual distance. The following cases can occur:
• Distance correct – meas. value correct -> basic setup completed
• Distance incorrect – meas. value incorrect -> a further interference echo mapping must be
carried out "check distance" (051).
• Distance correct – meas. value incorrect -> check "empty calibr" (005)
Note!
After the basic setup, an evaluation of the measurement with the aid of the envelope curve
("envelope curve" (0E) function group) is recommended (→ ä 57).
Return to group selection
After the interference echo suppression the basic setting is finished and the instrument jumps
automatically back into the group selection.