Commissioning
40
6.2.2
Key combination (display and operating module VU331)
Locking
Press S, O and F simultaneously.
Unlocking
If you try to change a parameter, the following appears:
Press S, O and F simultaneously. The "unlock parameter" (0A4) function appears.
Enter "2457". Now parameters can be changed.
Parameters affected
Locking by the key combination affects the following parameters:
• Parameters of the instrument specific blocks (Sensor Block, Diagnostic Block, Display Block)
• Endress+Hauser service parameters
6.2.3
Locking parameter
Locking
Enter a number other than "2457" into the "unlock parameter" (0A4) function.
(FOUNDATION Fieldbus: Diagnsotic Block, parameter PAROPERATIONMODE)
Unlocking
Enter "2457" into the "unlock parameter" (0A4) function.
(FOUNDATION Fieldbus: Diagnostic Block, parameter PAROPERATIONMODE)
Parameters affected
Locking by the locking parameter affects the following parameters:
• Parameters of the instrument specific blocks (Sensor Block, Diagnostic Block, Display Block)
• Endress+Hauser service parameters
Levelflex M FMP41C with FOUNDATION Fieldbus
L00-fmrxf0a4-20-00-00-de-001
Endress + Hauser
Need help?
Do you have a question about the Fieldbus Levelflex M FMP41C and is the answer not in the manual?
Questions and answers