Changing The Pump Tubing; Figure 8 Changing The Pump Tubing - M&C PSS Series Instruction Manual

Gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

Flexible tube, conveying belt, contact pulleys and contact springs are the only parts of the pump subject
to wear. They are simple to change.
Note

15.2.1 CHANGING THE PUMP TUBING

1 Conveying belt
4 contact pulley
Figure 8
Changing the pump tubing
For changing the pump tubing please proceed as follows:
1. Unplug the pump from the mains voltage. The device needs to be voltage free.
2.
Open tube connections at the pump;
3. Press conveying belt
4. Take away conveying belt
tube connectors;
5. Press the two contact pulleys
the contact springs have to be changed (see chapter 15.2.2);
6. Put the new tubing set
Note
30
Peristaltic pump is under pressure! Do not open housing!
A peristaltic pump might be part of a system, which is under pressure.
Check pressure before opening peristaltic pump, and adjust pressure to
atmosheric pressure.
If you send back the peristaltic pump to the M&C service for repair, please
let us know what kind of condensate has been pumped.
Before sending the pump back clean all parts from dangerous or highly
aggressive contaminants.
2 S-bolt
3 Tubing set
5 springs
at the recessed grips and turn S-bolt
and remove the old tubing set
and check whether the spring pressure is still sufficient, if not,
with the tube connectors into the guides of the conveying belt
Only the usage of the original tubing set guarantees a proper functionality.
Never lubricate the tube.
Before mounting the pump check all parts for contaminations and clean if
necessary.
SS-M05_Marine | 1.01.00
clockwise up to limit stop;
from the guides by pulling on the
www.mc-techgroup.com
;