15

MAINTENANCE

Maintenance work
Check, clean or replace of the filter
Check and replace of the filter
Check for condensate in the lines after the gas
cooler
Check the temperature of the Peltier sample
gas cooler
Replace hoses of the peristaltic pump
Replace valve plate and O-rings of the bellow
valve pump
Replace bellows of the below valve pump
M : M&C personnel or other trained personnel
1. Maintenance work must be carried out only by suitably trained specialist personnel.
2. The recommended maintenance intervals are based on M&C's experience.
3. The maintenance intervals can be further optimised if indicated by experience and local
conditions.
4. Depending on the composition of the measuring gas, it could be possible that maintenance
intervals have to be shortened. For costs resulting of this M&C will not take any responsibility.
Observe the plant- and process-specific safety measures prior to carrying out maintenance work!
W a r n i n g
N o t e
www.mc-techgroup.com
Dangerous voltage. before carrying out any work on the gas
conditioning unit, move the main switch to position '0' and pull out the
mains plug !
In order to protect downstream analyzers, the wet filter element must
always be replaced in the event of a condensate breakthrough.
SS-M05_Marine | 1.01.00
Maintenance intervals in
BMK
Option
1
-FI1
C
-FI2
-E1
-M2
-M1
-M1
C : Customer
months
3
6
12
36
C
C
C
C
M
M
27