Figure 5 Media Connection For Downstream Analyzer - M&C PSS Series Instruction Manual

Gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

Media connection for downstream analyzer
Figure 5
Media connection for downstream analyzer
The sample gas inlet is a Ø 6 mm tube.
All tubing connections are equipped with 4/6mm sealing ring threaded hose couplings made of PVDF,
for gas input temperatures of up to a maximum of 80 °C.
If heated sample lines are used, whereby the gas input temperatures are increased up to a maximum
of 180 °C, additional bulkhead unions made of stainless steel are necessary.
Connection tubing with 4/6 mm diameter are used for all models.
The sample gas and condensation tubing are to be assembled as follows:
Remove the union nut from the sealing ring couplings by turning it anti-clockwise. The nut should
be removed from the thread with great care to ensure that the loose sealing ring in the nut is not
lost.
Place the union nut over the connecting tubing.
Place the sealing ring over the connecting tubing with the thicker bead towards the nut.
Place the tubing over the nipple on the thread.
N o t e
Tighten the union nut by hand.
The tubing is now securely positioned and pressure-proof mounted.
Disassemble the tubing in reverse order.
www.mc-techgroup.com
The tightness of the connections can only be guaranteed if the
connecting tube has a straight rim (hose cutter).
Aggressive condensate possible!
Media residues in tubing!
Chemical burns caused by aggressive media possible!