Media connection for downstream analyzer
Figure 5
Media connection for downstream analyzer
The sample gas inlet is a Ø 6 mm tube.
All tubing connections are equipped with 4/6mm sealing ring threaded hose couplings made of PVDF,
for gas input temperatures of up to a maximum of 80 °C.
If heated sample lines are used, whereby the gas input temperatures are increased up to a maximum
of 180 °C, additional bulkhead unions made of stainless steel are necessary.
Connection tubing with 4/6 mm diameter are used for all models.
The sample gas and condensation tubing are to be assembled as follows:
Remove the union nut from the sealing ring couplings by turning it anti-clockwise. The nut should
•
be removed from the thread with great care to ensure that the loose sealing ring in the nut is not
lost.
Place the union nut over the connecting tubing.
•
Place the sealing ring over the connecting tubing with the thicker bead towards the nut.
•
Place the tubing over the nipple on the thread.
•
N o t e
Tighten the union nut by hand.
•
The tubing is now securely positioned and pressure-proof mounted.
Disassemble the tubing in reverse order.
www.mc-techgroup.com
The tightness of the connections can only be guaranteed if the
connecting tube has a straight rim (hose cutter).
Aggressive condensate possible!
Media residues in tubing!
Chemical burns caused by aggressive media possible!
SS-M05_Marine | 1.01.00
21