11. Make sure that the regulator valve is in the fully
closed position (Off) then disconnect the regulator
from the gas cylinder (turn regulator body in a counter
clockwise direction).
12. Replace the cap on the calibration gas cylinder.
13. Repeat steps 1 to 12 for each range to be calibrated
otherwise quit FCM. See QUITTING FCM for further
details.

Quitting FCM

Quit and Save Changes

1. Press and hold
2. Double press

Quit Without Saving Changes

1. Double press
Note: When you exit the FCM without saving the new
CAL data, the old calibration data remains in the
instrument memory.
B-8
FIRST RESPONDER CG - USER HANDBOOK
(Button 2) to save CAL data.
(Button 1) to exit FCM.
(Button 1) to exit FCM.