7. PERIODIC MAINTENANCE AND CLEANING

7.1. PERIODIC MAINTENANCE

 Regularly check the water level in the tank and add more if necessary.

7.2. CLEANING

 Prior to cleaning please unplug the device and clean the device at the room temperature.
 A piece of wet cloth may be used in order to remove the dust and dirt from the device.
 Mild detergent use is recommended to remove difficult dust and dirt.
 Please be aware of the undesirable effects of the chemicals and be careful while applying
them.
 Protect the device against rust. Please clean the rust as soon as possible when it is detected.

8. DISPOSAL MANAGEMENT CONCEPT

The user is responsible for the proper disposal of each individual component. All parts which
may comprise potentially infectious materials have to be disinfected by suitable validated
procedures (i.e. autoclaving, chemical treatment) prior to disposal.

9. TROUBLESHOOTING

If the device fails to operate, please check the followings:
 The power switch is on;
 The fuse is not blown;
 The plug is plugged-in properly;
 The plug is not defective;
 The installation of the plug is not defective;
 The mains supply is present.

9.1. ERROR CODES

In case of below written failures, related error codes are shown on the display.
OFL: The chamber temperature is higher than 100.5°C and/or
the temperature sensor endings
Switch off the device and contact to the authorized service.
are broken.
If an error occurs, please contact to an authorized Nüve agent to seek technical help.

9.2. FUSE REPLACEMENT

The fuses shall be always be replaced by the authorized personnel.