5.16. Replacing Locking Bobbin ( see Figure 7 & Figure 9)
Remove the flywheel handles (4).
•
Remove the lid protection cover (3) by removing its 4 screws.
•
Take the locking bobbin out by removing the bobbin (14) connection screws.
•
Separate the locking bobbin (14) electrical connection cables by cutting their welding points.
•
Take the pin on the locking bobbin and fix it onto the new bobbin.
•
Mount the new bobbin.
•
Connect the electrical connection cables by welding them carefully.
•
5.17. Replacing Locking Switch ( see figure 10)
Disconnect the locking switch (30) terminal connections.
•
Take the locking switch out by removing the locking switch connection nuts.
•
Mount the new locking switch and make the necessary electrical connections carefully according to
•
the electrical circuit diagram (figure 3 or Figure 4).
Adjust the switching length carefully.
•
5.18. Cleaning The Sterilizatior Chamber Filter ( see Figure 8 & Figure 11)
Take the shelf carrier (12) and the heater protection sheet (11) out by removing the shelf carrier
•
mounting screws (34) in the sterilization chamber.
Take the filter (36) out which is placed on the bottom of the chamber by removing the water inlet
•
record (35) with a wrench.
Clean the filter pores and re-mount it.
•
5.19. Replacing Pressure Switch ( see Figure 10)
Disconnect the cable connections of pressure switch terminals
•
Remove the pressure switch
•
Put some liquid gasket to female connectors of the new pressure switch to ensure that there is no
•
leakage and mount the new pressure switch
Make the electrical connections carefully according to the electrical circuit diagram (Figure 3 or
•
Figure 4)
Adjust the minimum pressure switch (33) by turning the red adjusting switch counter-clockwise or
•
adjust the maximum pressure switch (32) by turning the red adjusting switch clockwise.
17