NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. FN 300 - FN 400 - FN 500 FN 400 P – FN 500 P DRY AIR STERILIZERS/OVENS USER’S MANUAL NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. Z14.K 25 002 Rev. No:01 Rev. Date: 09/2002...
Page 2
Fax : (00 90 312) 399 21 97 Tel: 00 30 10 9028051-4 e-mail: Fax: 00 30 10 9029855 nuveexp@nuve.com.tr e-mail: info@arel.gr WARRANTY CERTIFICATE 1. Nüve warrants that the equipment delivered is free from defects in material and workmanship. This warranty is provided for a period of one year. The warranty period starts from the delivery date.
1. USE AND FUNCTION The FN series dry air sterilizers/ovens are designed to sterilize, heat and dry samples in laboratories. They offer excellent sterilization, drying and heating conditions. They maintain temperatures between 70 °C and 250 °C and keep the temperature stable within the given tolerances.
TECHNICAL SPECIFICATIONS 2.1. Technical Specifications Table FN 300 FN 400 FN 400 P FN 500 FN 500 P Temperature range 70°C / 250°C Temperature sensor Fe-Const Air forced ventilation Optional Optional Standard Optional Standard Control system PID program mable microprocessor Temperature set and 1°C...
INSTALLATION PROCEDURE 3.1. Lifting And Transport Because of the heavy weight of the dry air sterilizer/oven, all lifting and transport must be carried out using proper handling equipment. The dry air sterilizer/oven must be supported from underneath and never be turned over. 3.2.
Check the followings, ! the proposed site is suitable for users, ! the operator can follow up the unit even he deals with something else. ! the dry air sterilizer/oven does not occupy the utilisation space of others or damage them.
4.2. Control Thermostat 1. Temperature Display: This display shows the chamber temperature during the operation and the set temperature value during programming. The error codes are also shown on this display. 2. Time Display: This display shows the elapsed time during the operation and the set value during programming.
10. Alarm Led: It lights up when the program ends and if any failure occurs during the operation. 4.3.Programming Push the temperature set key. Set the temperature by pushing the value increase/decrease keys. Push the temperature set key again to save the temperature value.
±2 °C alarm limit. The sterilization operation is interrupted. You may re-start the dry air sterilizer/oven by pushing the start/stop key. PLEASE CONTACT TO AN AUTHORIZED NUVE AGENT TO SEEK TECHNICAL HELP IF AN ERROR OCCURS.